Name: | GE-MP portable coal gas analyzer |
---|---|
Measuring range: | 0~100% |
Resolution: | 0.01%/0.1% |
Accuracy: | ≤±2%F.S. |
Signal output: | 4~20mA or 0~5V |
Communication interface: | Rs-232 or Rs-485 Modbus RTU |
Power supply: | 100~240VAC 50/60Hz |
Measuring range: 0~100%
Resolution: 0.01%/0.1% (according to the measuring range selected)
Accuracy: ≤±2%F.S.
Repeatability: ≤±1%F.S.
Stability:
Zero drift: ±1%F.S./7d
Span drift: ±1%F.S./7d
Response time T90≤30s
Warm-up time: ≤45min
Sample gas flow rate: 1000±100mL/min
Sample gas pressure: 0.1MPa≤entrance pressure≤0.2MPa
Power supply: 100~240VAC 50/60Hz 1A
Ambient temperature: -5℃~+45℃
Environment humidity: ≤90%RH
Signal output: 4~20mA or 0~5V
Communication protocol: Rs-232 or Rs-485 Modbus RTU
Overall dimensions:
standard: 520 (W) x320 (H) x430 (D), (mm)
Weight: 10Kg
Measuring gas | Measuring range |
CO | 0-100% |
CO2 | 0-100% |
O2 | 0-100% |
CH4 | 0-100% |
H2 | 0-100% |
CnHm | 0-10% |
The GE-MP multi-component gas analyzer is a portable gas analyzer that is suitable for calorific value analysis of various gases.
Automatic calculation and display of the calorific value of the gas in units of large or megajoules
Automatic analysis and detection, simple operation, can replace gas analysis such as Austrian and chromatogram
The instrument is easy to carry, suitable for direct sampling analysis on industrial field pipelines, and sample analysis of laboratory airbags (bags)
RS-232 digital output interface with Modbus RTU communication protocol
The battery can be used for more than 6 hours on a single charge.
320*240 LCD monitors, desktop or panel mounted, easy to install
Imported infrared sensors, long service life, high sensitivity and fast response
Built-in precision filter, greatly improving the sensor life
Upper and lower limits of control points can be set at full scale range
Data backup, data recovery
Standard 4~20mA or 0~5V output
Standard Rs232 or Rs485 communication
Full English operation menu
Built-in precision air pump (optional)
The GE-MP coal gas analyzer uses analytical calculation methods to quickly and accurately analyze the concentration values of one or more gases and calculate the corresponding calorific value based on the concentration values.
The gas analyzer adopts imported infrared sensor, fuel cell oxygen sensor and thermal conductivity sensor, and uses advanced digital processing technology to realize continuous automatic and fast online detection of CO2, CO, CH4, H2, O2, CnHm and other gases.
Used for the detection of various industrial gases such as blast furnace gas, converter gas, coke oven gas, producer gas, water gas,
Calcium carbide furnace tail gas, natural gas, etc.
New energy industries such as bio-fermentation, bio-cracking, etc.
Determination of gas content in various combustion tests in the laboratory