Name: | PH8020 pure water pH electrode |
---|---|
Measuring range: | 0-14pH |
Temperature range: | 0-85℃ |
Working pressure: | 0.6MPa |
Measuring range: 0-14pH
Temperature range: 0-85℃
Slope: ≥96%
Zero point potential: E0=7PH±0.3
Pressure resistance: 0.6MPa
Installation size: upper and lower 3/4NPT pipe thread
Material quality: PPS/PC
Internal resistance: MΩ(25℃)<50MΩ
Connection method: low-noise cable leads directly
☆ Adopt international advanced solid dielectric and large area PTFE liquid junction. Not easy to block, easy to maintain
☆ The long-distance reference diffusion path greatly prolongs the service life of the electrode in harsh environments
☆ The newly designed glass bulb increases the bulb area and prevents the formation of interfering bubbles in the internal buffer, making the measurement more reliable
☆ Adopt PPS/PC shell, upper and lower 3/4NPT pipe thread, easy to install, no need for sheath, saving installation cost
☆ The electrode adopts high-quality and low-noise cables, which can make the signal output longer than 20 meters without interference
☆ The electrode is made of ultra-bottom impedance sensitive glass film, which has the characteristics of fast response, accurate measurement, good stability and not easy to hydrolyze in the occasion of low conductivity and high purity water.
☆ High measurement accuracy, fast response and good repeatability
☆ Ag/AgCL reference electrode with silver ion
☆ Correct operation and long service life
☆ Can be installed sideways or vertically in the reaction tank or pipeline
☆ The electrode can be used interchangeably with similar foreign electrodes
Basic characteristics of PH pure water electrode
1. Filled with high molecular polymer, the reference junction potential is very stable.
2. The diffusion potential is very stable. Because the large-area diaphragm is distributed around the glass bubble, the distance between the reference diaphragm and the glass bubble is constant and close, and the ions diffused by the diaphragm quickly form a complete measurement with the glass electrode. The loop responds quickly, so that the diffusion potential is not easily affected by the external flow rate and so on, so it is very stable.
3. Since the diaphragm is filled with high molecular polymer, the electrolyte outflow is small and stable, and will not pollute the pure water measured.
Therefore, the above characteristics of the composite electrode make it very suitable for measuring the PH value of high-purity water!
Measurement of pure water and high-purity water in pharmaceutical, chlor-alkali chemicals, pigments and dyes, pulp and paper, intermediates, fertilizers, starch, environmental protection and other industries