Name: | DCC1 RGline online residual chlorine electrode for swimming pool |
---|---|
Measuring range: | 0.005~2.00ppm 0.05~5.00/10.00/20.00ppm |
Working temperature: | 0~45℃ |
Working pressure: | 0.5bar |
Output: | 4-20mA、Modbus485、0~±2V |
Measured medium | Free chlorine |
Measuring principle | Coated three-electrode current method |
Measuring range | 0.005~2.00ppm 0.05~5.00/10.00/20.00ppm |
pH range | 4.0~12.0 |
Working temperature | 0~45℃ |
Flow speed | 15~30L/H |
Working pressure | 0.5bar |
Zero calibration | Not needed |
Slope calibration | According to DPD-1 method, calibrate on the device |
Drift | <-3%/month |
Response time | It takes 2 minutes to reach 90% of the measured value |
Polarization time | 2 hours |
Electrical connection | 4-pole plug, 5-pole M12 connector, 2-pole terminal |
Power supply | DC 12~30V, ±5~±15V |
Signal output | 2-wire 4-20mA, Modbus485, 0~±2V |
Dimensions | Diameter: 25mm Length: 175~220mm |
Material | Microporous hydrophilic membrane, PVC-U, PEEK, stainless steel |
Maintenance interval | Membrane cap: replace once a year Electrolyte: replace 3~6 months |
Storage | Electrode rod and membrane cap: 5~40℃ dry environment Electrolyte: 5~35℃ in a cool and dark place |
CE | EMC, RoHS |
Pool water containing cyanuric acid, electrolysis produces chlorine or chloride
Drinking water, sewage and sea water used for disinfection.
Pool water containing cyanuric acid, electrolysis produces chlorine or chloride
Drinking water, sewage and sea water used for disinfection.
Pool water containing cyanuric acid, electrolysis produces chlorine or chloride
Drinking water, sewage and sea water used for disinfection.