PRODUCT CENTER
Name:Online turbidity sensor
Measuring range:0.01-100 NTU, 0.01-4000 NTU
Accuracy:<±1%F.S.
Working temperature:0~45℃

Item

Parameters

Power supply

12VDC

Dimensions

diameter 60mm * length 256mm

Weight

1.65Kg

Material

Body:SUS316L + PVC (normal version), titanium alloy (sea water version)

O-ring: fluororubber

Cable: PVC

Protection

IP68/NEMA6P

Measuring range

0.01-100 NTU, 0.01-4000 NTU

Accuracy

<±1%F.S.

Working pressure

≤0.4Mpa

Flow speed

≤2.5m/s, 8.2ft/s

Working temperature

0~45℃

Calibration

Sample calibration, slope calibration

Cable

Standard 10 meters cable, can be extended to 100 meters

Warranty

1 year


The sensor is based on the combined infrared absorption and scattered light method, and the ISO7027 method can be used to continuously and accurately measure turbidity. The turbidity value is determined according to the ISO7027 infrared double-scattering light technology without being affected by chromaticity. The self-cleaning function can be selected according to the use environment. Stable data and reliable performance; built-in self-diagnosis function to ensure accurate data; simple installation and calibration. It can be widely used in turbidity monitoring in sewage plants, water plants, water stations, surface water, industry and other fields.

The sensor is based on the combined infrared absorption and scattered light method, and the ISO7027 method can be used to continuously and accurately measure turbidity. The turbidity value is determined according to the ISO7027 infrared double-scattering light technology without being affected by chromaticity. The self-cleaning function can be selected according to the use environment. Stable data and reliable performance; built-in self-diagnosis function to ensure accurate data; simple installation and calibration. It can be widely used in turbidity monitoring in sewage plants, water plants, water stations, surface water, industry and other fields.

It can be widely used in turbidity monitoring in sewage plants, water plants, water stations, surface water, industry and other fields.